1.
Anal Bioanal Chem ; 415(5): 809-821, 2023 Feb.
Artículo en Inglés | MEDLINE | ID: mdl-36507958

RESUMEN

The US National Institute of Standards and Technology (NIST) developed a Standard Reference Material® (SRM®) 3949 Folate Vitamers in Frozen Human Serum to replace SRM 1955 Homocysteine and Folate in Human Serum. The presence of increased endogenous levels of folic acid and 5-methyltetrahydrofolate (5mTHF) in SRM 3949, enhanced folate stability via addition of ascorbic acid, and inclusion of values for additional minor folates are improvements over SRM 1955 that should better serve the clinical folate measurement community. The new SRM contains folates at three levels. To produce SRM 3949, pilot sera were collected from 15 individual donors, 5 of whom were given a 400-µg folic acid supplement 1 h prior to blood draw to increase serum levels of 5mTHF and folic acid for the high-level material. To stabilize the folates, 0.5% (mass concentration) ascorbic acid was added as soon as possible after preparation of serum. These pilot sera were screened for five folates plus the pyrazino-s-triazine derivative of 4-α-hydroxy-5-methyltetrahydrofolate (MeFox) at the US Centers for Disease Control and Prevention (CDC) by isotope dilution liquid chromatography-tandem mass spectrometry (ID-LC-MS/MS). Based on these results, a blending protocol was specified to obtain the three desired folate concentrations for SRM 3949. ID-LC-MS/MS analysis at the CDC and NIST was utilized to assign values for folic acid and 5mTHF, as well as several minor folates.


Asunto(s)
Ácido Fólico , Espectrometría de Masas en Tándem , Humanos , Ácido Fólico/análisis , Cromatografía Liquida/métodos , Espectrometría de Masas en Tándem/métodos , Estándares de Referencia , Ácido Ascórbico
2.
Anal Chem ; 89(19): 10461-10467, 2017 10 03.
Artículo en Inglés | MEDLINE | ID: mdl-28930436

RESUMEN

A new tobacco filler Standard Reference Material (SRM) has been issued by the National Institute of Standards and Technology (NIST) in September 2016 with certified and reference mass fraction values for nicotine, N-nitrosonornicotine, 4-(methylnitrosamino)-1-(3-pyridyl)-1-butanone, and volatiles. The constituents have been determined by multiple analytical methods with measurements at NIST and at the Centers for Disease Control and Prevention, and with confirmatory measurements by commercial laboratories. This effort highlights the development of the first SRM for reduced nicotine and reduced tobacco-specific nitrosamines with certified values for composition.


Asunto(s)
Cromatografía de Gases y Espectrometría de Masas/métodos , Espectrometría de Masas en Tándem/métodos , Productos de Tabaco/análisis , Cromatografía Líquida de Alta Presión/métodos , Cromatografía Líquida de Alta Presión/normas , Congelación , Cromatografía de Gases y Espectrometría de Masas/normas , Nicotina/análisis , Nicotina/normas , Nitrosaminas/análisis , Nitrosaminas/normas , Transición de Fase , Estándares de Referencia , Espectrometría de Masas en Tándem/normas , Productos de Tabaco/normas , Compuestos Orgánicos Volátiles/análisis , Compuestos Orgánicos Volátiles/normas